Comparative analysis of deformable mirrors for ocular adaptive optics

Open Access Open Access

Abstract

We have evaluated the ability of three commercially available deformable mirrors to compensate the aberrations of the eye using a model for aberrations developed by Thibos, Bradley and Hong. The mirrors evaluated were a 37 actuator membrane mirror and 19 actuator piezo mirror (OKO Technologies) and a 35 actuator bimorph mirror (AOptix Inc). For each mirror, Zernike polynomials and typical ocular aberrated wavefronts were fitted with the mirror modes measured using a Twyman-Green interferometer. The bimorph mirror showed the lowest root mean square error, although the 19 actuator piezo device showed promise if extended to more actuators. The methodology can be used to evaluate new deformable mirrors as they become available.

©2005 Optical Society of America

Full Article  |  PDF Article
More Like This
Correction of ocular and atmospheric wavefronts: a comparison of the performance of various deformable mirrors

Nicholas Devaney, Eugenie Dalimier, Thomas Farrell, Derek Coburn, Ruth Mackey, David Mackey, Francois Laurent, Elizabeth Daly, and Chris Dainty
Appl. Opt. 47(35) 6550-6562 (2008)

Adaptive optics with a magnetic deformable mirror: applications in the human eye

Enrique J. Fernández, Laurent Vabre, Boris Hermann, Angelika Unterhuber, Boris Považay, and Wolfgang Drexler
Opt. Express 14(20) 8900-8917 (2006)

Membrane deformable mirror for adaptive optics: performance limits in visual optics

Enrique J. Fernández and Pablo Artal
Opt. Express 11(9) 1056-1069 (2003)

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (11)

Fig. 1.
Fig. 1. Actuators layout for the three mirrors considered: the 37ch OKO MMDM, the 19ch OKO, and the 35ch Bimorph. The gray area represents the optical pupil typically used. It is defined as such by the manufacturer for the 19 actuator OKO and the 35 actuator AOptix mirrors, but set by the user for the 37 actuator OKO mirror.
Fig. 2.
Fig. 2. Schematic of the 37ch OKO mirror actuation and examples of actuator response.
Fig. 3.
Fig. 3. Schematic of the 19ch OKO mirror actuation and examples of actuator response.
Fig. 4.
Fig. 4. Measurement of hysteresis through the ramping of one actuator.
Fig. 5.
Fig. 5. Schematic of the 35ch AOptix mirror actuation and examples of actuator response.
Fig. 6.
Fig. 6. Comparison of the Zernike generation for the three mirrors in terms of pv signed wavefront produced. The first 21 polynomials are represented (piston, tip and tilt were removed).
Fig. 7.
Fig. 7. Comparison of the Zernike generation for the three mirrors: (a) in terms of maximum signed wavefront rms and (b) in terms of residual rms error.
Fig. 8.
Fig. 8. Residual wavefront rms error after fitting with the three mirrors over a 6 mm pupil. Piston, tip and tilt terms were removed
Fig. 9.
Fig. 9. Residual rms wavefront error after fitting with the three mirrors. The results were averaged from calculations on the same initial wavefronts as in Fig. 8.
Fig. 10.
Fig. 10. Comparison of residual rms obtained with an open-loop correction and a closedloop correction with the 37 actuator OKO MMDM
Fig. 11.
Fig. 11. Closed-loop correction of a human eye with the AOptix mirror (pupil size for measurement: 4.8 mm).

Equations (7)

Equations on this page are rendered with MathJax. Learn more.

M = U M V T
Φ M = U U T Φ
c = M + Φ = V W 1 U T Φ
Z m = c lim c max U U T Z
c max = max ( abs ( M + Z ) )
Φ M = U W V T f ( V W 1 U T Φ )
f ( c i ) = { c if c lim c i c lim c lim × c i abs ( c i ) if c i < c lim or c i > c lim