Optical coherence tomography using a continuous-wave, high-power, Raman continuum light source

Open Access Open Access

Abstract

High performance, short coherence length light sources with broad bandwidths and high output powers are critical for high-speed, ultrahigh resolution OCT imaging. We demonstrate a new, high performance light source for ultrahigh resolution OCT. Bandwidths of 140 nm at 1300 nm center wavelength with high output powers of 330 mW are generated by an all-fiber Raman light source based on a continuous-wave Yb-fiber laser-pumped microstructure fiber. The light source is compact, robust, turnkey and requires no optical alignment. In vivo, ultrahigh resolution, high-speed, time domain OCT imaging with <5 µm axial resolution is demonstrated.

©2004 Optical Society of America

Full Article  |  PDF Article
More Like This
Ultrahigh resolution optical coherence tomography imaging with a broadband superluminescent diode light source

Tony H. Ko, Desmond C. Adler, James G. Fujimoto, Dmitry Mamedov, Viatcheslav Prokhorov, Vladimir Shidlovski, and Sergei Yakubovich
Opt. Express 12(10) 2112-2119 (2004)

Ultrahigh resolution real time OCT imaging using a compact femtosecond Nd:Glass laser and nonlinear fiber

S. Bourquin, A. D. Aguirre, I. Hartl, P. Hsiung, T. H. Ko, J. G. Fujimoto, T. A. Birks, W. J. Wadsworth, U. Bünting, and D. Kopf
Opt. Express 11(24) 3290-3297 (2003)

Continuum generation in a novel photonic crystal fiber for ultrahigh resolution optical coherence tomography at 800 nm and 1300 nm

Aaron D. Aguirre, Norihiko Nishizawa, James G. Fujimoto, Wolfgang Seitz, Max Lederer, and Daniel Kopf
Opt. Express 14(3) 1145-1160 (2006)

Supplementary Material (12)

Media 1: MPG (2357 KB)     
Media 2: AVI (8328 KB)     
Media 3: AVI (2371 KB)     
Media 4: AVI (6196 KB)     
Media 5: AVI (2336 KB)     
Media 6: AVI (8306 KB)     
Media 7: AVI (6196 KB)     
Media 8: AVI (1747 KB)     
Media 9: AVI (1747 KB)     
Media 10: MPG (2296 KB)     
Media 11: AVI (6024 KB)     
Media 12: MPG (2167 KB)     

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (9)

Fig. 1.
Fig. 1. Left: Schematic of Raman continuum source. SMF: Single-mode fiber. Right: Output spectrum of the Raman continuum source. The output was 5.5W total, with 2.3 W in the spectral range from 1090 to 1370 nm. The output was filtered using a special WDM coupler to remove the pump wavelength and shape the spectrum.
Fig. 2.
Fig. 2. (a) Typical output spectrum of the Raman fiber light source before and after spectral shaping. (b) RF noise spectrum. High frequency oscillations are caused by spurious reflections from the fiber splices. Dual-balanced detection is used to reduce excess noise.
Fig. 3.
Fig. 3. Schematic of high-speed, ultrahigh resolution OCT system using an all-fiber Raman continuum light source. The system uses broadband 80/20 and 50/50 couplers to optimize the power on the sample and coupled back to the detectors. Dispersion was matched in the sample and reference arms of the interferometer in order to maintain high axial resolution.
Fig. 4.
Fig. 4. (a) Point spread function measured using an isolated 3.0 OD attenuated reflection from a mirror. The axial resolution is 6.3 µm in air, corresponding to <5 µm in tissue. (b) Logarithmically demodulated signal showing low sidelobes.
Fig. 5.
Fig. 5. In vivo high-speed OCT images of human skin. The stratum corneum layer (SC), epithelium (E), dermis, dermal-epidermal junction (inset left) and a spiraling sweat duct (inset right) can clearly be seen. Images were acquired at 3.2 frames per second. (~18 µm×4.8 µm transverse x axial resolution; 500×1000 pixels; 2.25 mm×1.8 mm).
Fig. 6.
Fig. 6. High-speed, ultrahigh resolution in vivo OCT image of Syrian hamster cheek pouch. The keratinized epithelial layer (e), muscular layers (m), and two prominent vessels (v) are clearly visible. A microscope cover glass (c) was placed over the top of the cheek pouch. (~18 µm×4.8 µm transverse x axial resolution; 500×1000 pixels; 2.4 mm×1.4 mm).
Fig. 7.
Fig. 7. Three-dimensional volume imaging using in vivo ultrahigh resolution OCT images acquired at 3.2 frames per second. (a). (2.3 MB) Movie showing sequential transverse slices of hamster cheek pouch with two large vessels, corresponding to the normal OCT view (6.0 MB version). (b) (2.4 MB) Movie showing en face slices at different depths, perpendicular to the OCT view, allowing features such as epithelial folds and vessels to be visualized (8.3 MB version). The rendered volumes cover 2.4 mm×1.0 mm×1.8 mm.
Fig. 8.
Fig. 8. (2.4 MB) Movie showing rendered volume of hamster cheek pouch constructed from a three-dimensional dataset. The tissue volume can be viewed from arbitrary virtual perspectives. The rendered volume covers 2.4 mm×1.0 mm by approximately 1.8 mm.
Fig. 9.
Fig. 9. (a) (2.2 MB) Movie showing en face slices of human fingerpad created from cross-sectional images. Spiraling sweat ducts are seen along the fingerpad ridges (2.25×0.75×1.8 mm) (8.3 MB version). (b) (2.3 MB) Movie showing 3D rendering and segmentation of sweat ducts (6.2 MB version). [Media 7] Duct density as well as individual ducts can be assessed (1.7 MB). [Media 9]
Select as filters


Select Topics Cancel
© Copyright 2024 | Optica Publishing Group. All rights reserved, including rights for text and data mining and training of artificial technologies or similar technologies.