Ultrahigh resolution optical coherence tomography using a superluminescent light source

Open Access Open Access

Abstract

A superluminescent Ti:Al2O3 crystal is demonstrated as a light source for ultrahigh resolution optical coherence tomography (OCT). Single spatial mode, fiber coupled output powers of ~40 μW can be generated with 138 nm bandwidth using a 5 W frequency doubled, diode pumped laser, pumping a thin Ti:Al2O3 crystal. Ultrahigh resolution OCT imaging is demonstrated with 2.2 μm axial resolution in air, or 1.7 μm in tissue, with >86 dB sensitivity. This light source provides a simple and robust alternative to femtosecond lasers for ultrahigh resolution OCT imaging.

©2002 Optical Society of America

Full Article  |  PDF Article
More Like This
Ultrahigh resolution optical coherence tomography imaging with a broadband superluminescent diode light source

Tony H. Ko, Desmond C. Adler, James G. Fujimoto, Dmitry Mamedov, Viatcheslav Prokhorov, Vladimir Shidlovski, and Sergei Yakubovich
Opt. Express 12(10) 2112-2119 (2004)

Ultrahigh-resolution optical coherence tomography with a diode-pumped broadband Cr3+:LiCAF laser

Philipp C. Wagenblast, Tony H. Ko, James G. Fujimoto, Franz X. Kaertner, and Uwe Morgner
Opt. Express 12(14) 3257-3263 (2004)

Ultrahigh-resolution optical coherence tomography at 1.15 μm using photonic crystal fiber with no zero-dispersion wavelengths

Hui Wang, Christine P. Fleming, and Andrew M. Rollins
Opt. Express 15(6) 3085-3092 (2007)

Cited By

Optica participates in Crossref's Cited-By Linking service. Citing articles from Optica Publishing Group journals and other participating publishers are listed here.

Alert me when this article is cited.


Figures (5)

Figure 1.
Figure 1. Experimental schematic of low coherence light source. A high doping density Ti:Al2O3 crystal enables pumping with a high excitation density per unit volume. A double pass pump and fluorescence collection geometry is used.
Figure 2.
Figure 2. Output power coupled into a single mode optical fiber versus pump power. A double pass configuration approximately doubles the collected power.
Figure 3.
Figure 3. (left) Spectrum after coupling into a single mode optical fiber. The bandwidth is 138 nm centered at 761 nm. (right) Interference fringes from photodetector at the output of the OCT dual balanced interferometer. The axial resolution is 2.2 μm in air, corresponding to 1.7 μm in tissue.
Figure 4.
Figure 4. Schematic of ultrahigh resolution OCT system. A dual balanced interferometer was used to cancel the excess noise in the light source. Dispersion was balanced in the sample and reference arms.
Figure 5.
Figure 5. Ultrahigh resolution OCT imaging of African Frog (Xenopus Laevis) tadpole in vivo. (Left) Imaging was performed with <2 μm axial resolution and 5 μm transverse resolution. (Right) A composite image with extended depth of field formed by fusing 5 images with different focal depth settings.